www.cs-instruments.com
4
Compressed air analysis
with DS 300
The multi function display DS 300 gives the user a survey on the whole
compressed air net
Graphic evaluation of all
relevant data
All relevant measuring data for an
economic operation of the compressed air
line at a glance:
• Actual air consumption in m³/h
or m³/min
• Total air consumption in m³
• Pressure dew point in °Ctd
• Line pressure in bar
• Current consumption in A
• Temperature in °C
The data transfer to the PC is effected:
• by means of the CS data collector
(manually)
• via USB interface for short distance PC
to display DS 300
• via RS 485 interface for long distances
The CS data collector (USB) is able to:
• read out up to 64 x DS 300 at the PC
• evaluate data in the CS Soft Professional
• reprogram DS 300
TECHNICAL DATA
Consumption probe:
please see page 6
Multifunction display DS 300:
please see page 2
CLAMP-ON AMMETER:
Measuring range: 0 to 500 A
Accuracy: ± 0.5 %
DEW POINT PROBE FA 300-2:
Measuring range: -80 to 20 °Ctd
Accuracy: ± 2 °Ctd at -40 °Ctd
PRESSURE PROBE:
Measuring range: 0 to 16 bar
Accuracy: ± 0,5 %
USB
CS-BUS
(RS-485)
Alarm Relay
4 ... 20 mA output
bar